8 General technical specifications Power supply 90W power supply unit Input range: 100–240V, 50–60Hz, 115V, 400Hz nominal for aircraft use Typical
English 9 VPAP ST technical specifications Mode pressure ranges CPAP mode Set Pressure: 4–20 cm H2O S, ST, T and PAC modes IPAP: 4–25 cm H2O; EPAP
10 Air tubing technical specifications Air tubing Material Length Inner diameter ClimateLine heated air tubing Flexible plastic and electrical
English 11 Humidifier performance The following settings have been tested at 22°C ambient temperature: Mask pressure cm H2O RH output % Nominal sy
12 Displayed values Value Range Display resolution Pressure sensor at air outlet Mask pressure 4-20 cm H2O (CPAP); 2-25 cm H2O (S, ST, T, PAC, i
English 13 Pressure accuracy Maximum static pressure variation at 10 cm H2O according to EN ISO 17510-1:2009 Standard air tubing SlimLine air tub
14 Direct current; Lock/unlock; China pollution control logo 1; China pollution control logo 2; European Authorised Representative; Not d
English 15 Product Warranty period Clip-type finger pulse sensors CPAP and bilevel device data modules Oximeters and CPAP and bilevel de
16 WARNINGS Read the entire manual before using the device. Use the device only as directed by your physician or healthcare provider. Use
English 17 Do not leave long lengths of air tubing around the top of your bed. It could twist around your head or neck while you are sleeping.
Making quality of care easyRespiratory Care Solutions368551.indb 2 4/02/2013 3:14:20 PM
Making quality of care easyRespiratory Care Solutions368551.indb 3 4/02/2013 3:14:20 PM
Global leaders in sleep and respiratory medicine www.resmed.com Manufacturer: ResMed Ltd 1 Elizabeth Macarthur Drive Bella Vista NSW 2153 Australia
English 1 English Indications for use The VPAP ST is indicated to provide noninvasive ventilation for patients weighing more than 13 kg or more th
2 Setup Refer to illustration A. 1. Connect the DC plug of the power supply unit to the rear of the device. 2. Connect the power cord to the power
English 3 Notes: If your clinician has enabled SmartStart your device will start automatically when you breathe into the mask and stop automati
4 SD card An SD card has been supplied to gather therapy data from your device and provide settings updates from your clinician. When instructed to
English 5 Problem/Possible cause Solution EPAP may be set too low. Talk to your clinician about your settings. Device does not start when you br
6 Problem/Possible cause Solution The patient is using a paediatric mask with a set pressure less than 8 cm H2O. Disable SmartStart/Stop. Pressure
English 7 Problem/Possible cause Solution Displays message: No tube, please check your tube is connected Flow is high because air tubing is not c